What is the difference between calibration and verification?
Calibration establishes the relationship between indicated and reference values with a stated uncertainty. Performance verification confirms the machine still meets a stated specification, usually ISO 10360-2 or ASME B89.4.1. Most work sold as "CMM calibration" is the latter.
Do you offer CMM calibration near me?
On-site service covers Chicago and the listed metro area pages within the region. Outside that radius, machines are typically handled ship-in, with crating guidance provided in the quotation.
What happens if the machine fails as-found?
As-found deviations are recorded before adjustment and reported with the certificate, supporting a reverse-traceability impact assessment back to the last passing calibration.
What does ISO 10360-2 actually test?
Length measurement error, E0 and E150, is measured at five calibrated lengths, three times each, in seven positions, giving 105 measurements checked against the machine's stated maximum permissible error.
What temperature does the room need to be?
ISO 1 sets 20 °C as the dimensional reference temperature. Acceptance tests follow the environment the machine manufacturer specified; reverification conditions are whatever the user records at the time, which is why they belong on the certificate.
Is the certificate accredited, and how do I check the scope?
Certificates are issued under the performing laboratory's scope of accreditation, which can be verified in the accrediting body's public directory against the measurement, range and uncertainty quoted.
What is a CMC and why does it matter to my tolerances?
Calibration and measurement capability is the smallest uncertainty a laboratory can claim within its scope. If the CMC doesn't support a 4:1 test uncertainty ratio against your tolerance, the result may not be defensible to an auditor.
What is an error map, and can it be updated on my controller?
The volumetric compensation map corrects geometric error across the measuring volume. It's updated during geometric adjustment on controllers that allow write access; some OEM-locked controllers restrict this to the manufacturer.
Will this satisfy my IATF or AS9100 auditor?
Auditors under IATF 16949 clause 7.1.5.3 and AS9100's 7.1.5.2 look for a stated scope, traceability to the SI, and evidence the external laboratory is ISO/IEC 17025 accredited or otherwise acceptable to the customer. All of that appears on the certificate.
Can calibration be done on-site, and how long is the machine down?
Bridge and gantry machines are usually serviced on-site. Time on site and scheduling around shutdown windows are confirmed in the quotation for the specific machine and scope.

Further Detail
Standards, traceability and equipment control are set out on the quality system page. Competence and authorization are covered on the our team page. Commercial terms are in the terms and conditions.
Question about a specific item?
Send the detail and it is answered against the method that would actually be used.