What a calibration visit actually tests

Length measurement error (E0, E150) is verified per ISO 10360-2:2009 and probing error (PFTU) per ISO 10360-5, each checked against the machine's stated MPE rather than a generic tolerance.

Articulated arms are tested under ISO 10360-12, where applicable to the arm and probe configuration; the arm standard is never treated as an extension of a bridge-machine certificate. Laser tracker frames are tested under ISO 10360-10, with ASME B89.4.19 as the US analogue, and multisensor machines fall under ISO 10360-9.

Sequence of a typical visit

The order below follows how a technician actually works through a machine, not a sales pitch about outcomes.

  • 01

    Scales and reader head

    X, Y and Z scales are cleaned and inspected; reader head output is checked and adjusted where the controller permits.

  • 02

    Air bearings and airlines

    Bearing clearance and supply air quality are checked; contaminated air is a common cause of drift.

  • 03

    Geometry

    Straightness, linearity and squareness in the XY, ZX and ZY planes are measured with a laser interferometer or by sequential multilateration.

  • 04

    Length measurement error

    Five calibrated lengths are measured three times each in seven positions: 105 measurements against E0,MPE and E150,MPE.

As-found to as-left

  • 01

    As-found test

    Deviations are recorded against the stated MPE before any adjustment is made.

  • 02

    Geometric adjustment

    Kinematic errors are corrected and, where the controller allows it, the volumetric compensation map is updated.

  • 03

    As-left verification

    The E0/E150 and probing error tests are repeated to confirm the machine now meets specification.

  • 04

    Certificate issue

    Results, uncertainty, decision rule and traceability statement are issued under the performing laboratory's scope.

Why CMM Calibration and Repair Chicago

  • 01

    Repair, not just testing

    Probe heads, scales, reader heads, air bearings, drives and counterbalance systems are serviced on the same visit when needed.

  • 02

    OEM-locked maps respected

    Where a controller restricts compensation-map access to the OEM, that limit is stated plainly on the report.

  • 03

    Crash recovery and relocation

    Machines moved or struck are re-verified before being returned to production use.

Coverage across Illinois and the Greater Midwest

On-site scheduling covers the Chicago metro and the listed service areas. Machines outside that radius are quoted on a ship-in basis.

Discuss CMM Calibration and Repair Requirements

Describe the machine and the standard required for the certificate.

Pricing is set out in writing and sent by email.
Request a Quote + CertificationEmail us